Proceedings:
Applications of Computer Vision in Medical Image Processing
Volume
Issue:
Papers from the 1994 AAAI Spring Symposium
Track:
Contents
Downloads:
Abstract:
In this work, we present results from a new formulation for determining cross sectional blood velocities from a time.sequence of X-ray projection arteriograms. Starting with the conservation of mass principle, and physics of X-ray projection, we derive a motion constraint equation for projection imaging, a practical special case of which is shown to be the Horn and Schunck’s optical flow constrainL We are interested in the study of nonrigid motion of blood which is an incompressible fluid, and as such have developed a formulation for optical flow which is applicable to such media. The formulation is particularly efficient, as the flow field is obtained from a 90 degrees rotation applied to the gradient of a scalar function. It is shown that if specific criteria are met, in addition to normal flow which is commonly recoverable, the tangential component of flow is also recover. able, without the need for smoothness. An algorithm is presented to illustrate this. Preliminary results from the optical flow formulation applied to synthetic images, as well as contrast-injected X-ray images of flowing fluid, in a cylindrical vessel phantom are presented.
Spring
Papers from the 1994 AAAI Spring Symposium